Real-world Accuracy of CGM in Inpatient Critical and Noncritical Care Settings at a Safety-Net Hospital

OBJECTIVE
We sought to determine real-world accuracy of inpatient continuous glucose monitoring (CGM) at multiple levels of acuity in a large safety-net hospital.
RESEARCH DESIGN AND METHODS
We analyzed records from hospitalized patients on Dexcom G6 CGM, including clinical, point of care (POC), and laboratory (Lab) glucose, and CGM data. POC/Lab values were matched to the closest timed CGM value. Encounters were divided into not critically ill (NCI) versus critically ill (CI). CGM accuracy was evaluated.
RESULTS
Paired readings (2,744 POC-CGM; 3,705 Lab-CGM) were analyzed for 233 patients with 239 encounters (83 NCI, 156 CI). POC-CGM aggregated and average mean absolute relative differences (MARD) were 15.1% and 17.1%. Lab-CGM aggregated and average MARDs were 11.4% and 12.2%. Accuracy for POC-CGM and Lab-CGM was 96.5% and 99.1% in Clarke Error Grid zones A/B.
CONCLUSIONS
Real-world accuracy of inpatient CGM is acceptable for NCI and CI patients. Further exploration of conditions associated with lower CGM accuracy in real-world settings is warranted.